| FLAGS | What it means |
| Biomarker 2x of URR | Biomarker concentration for specific barcode is 2x higher than upper reference range |
| Biomarker 10x of URR | Biomarker concentration for specific barcode is 10x higher than lower reference range |
| Biomarker 50% of LRR | Biomarker concentration for specific barcode is 50% of lower reference range |
| Biomarker 10% of LRR | Biomarker concentration for specific barcode is 10% of lower reference range |
| Biomarker above toxic level | Biomarker concentration for specific barcode is above toxic level |
| Barcode hemolysed | Barcode may be hemolysed |
| Biomarker exposed to light | Vitamin B1, B2, B3-amide, B5, B6, B7, and C are all out-of-range (yellow/red) |
| Unusual yellow/red flags for biomarker | Compare % yellow/red for each biomarker this batch vs historical and flag depending on batch size BATCH SIZE 1-10: flag if biomarker shows 50% difference from historical BATCH SIZE 11-20: flag if biomarker shows 40% difference from historical BATCH SIZE 21-30: flag if biomarker shows 30% difference from historical BATCH SIZE >30: flag if biomarker shows 20% difference from historical |
| Barcode contains many outlier biomarkers | Flag samples with outlying biomarker % higher than 2SD (of all samples); samples with >30% outlier biomarkers excluded from SD calculation |
| Barcode contains many lower outlier biomarkers | Barcode has >=5 lower outliers & accounts for >=70% of all outliers |
| Percentage of outlying biomarkers >3SD for this barcode | Sample level rerun request: rerun samples with outlying biomarker % higher than 3SD (of all samples) |
| Key biomarker (serum albumin) out of range | Serum albumin has >4% as yellow/red this batch |
| Key biomarker (CRP) out of range | CRP has >10% as yellow/red this batch |
| Unusual LOD for biomarker | METAB/EXPO: biomarker is LOD & historically > 8% LOD & not NR PROTEIN: same as above + LOD value of this batch is less or equal to last batch |
| NIST concentration outside of 15% avg + biomarker flagged | Flag biomarkers (batch-wide) outside of +-15% average/NIST theoretical & if the biomarker has too many yellow/red flags this batch (theoretical concentration prioritized if available) |
| NIST concentration outside of 20% avg | Flag biomarkers (batch-wide) outside of +-20% average/NIST theoretical (theoretical concentration prioritized if available) |
| Pooled plasma outside of 20% avg + biomarker flagged | Flag biomarkers (batch-wide) outside of +-20% pooled plasma average & if the biomarker has too many yellow/red flags in this batch (combine with Row 9). |
| Pooled plasma outside of 30% avg | Flag biomarkers (batch-wide) outside of +-30% pooled plasma average |
| Control concentration changing in one direction | Flag if last 4 control concentration batches are going in the same direction & degree of change >2SD. |
| >20 biomarkers exceed the 30% CV in pooled plasma | If >20 biomarkers exceed the 30% CV, flag the batch as potential failure for rerun. |